INSTALLATION OF AIR/WATER SYRINGE KIT

Air/Water Syringe Repair Kit (General)
Product #: A113-019

Air/ Water Syringe Repair Kit (Euro)
Product #: A113-403

(Tools needed: Small Allen Key or a Hex Key)
Note: Petroleum Based Lubricant is needed for this repair

 

REMOVAL & INSTALLATION (GENERA – SQUARE HEADED SYRINGE)

  1. DEPRESSURIZE dental station.
  2. REMOVE PIN from the side of the SRINGE HOUSING by using a small PAPER CLIP or small ALLEN WRENCH.
  3. REMOVE BOTH BUTTONS by PULLING straight OUT.
  4. REMOVE BOTH VALVE CORES by UNSCREWING with tool that comes with the kit.
  5. CLEAN BOTH BUTTON HOLES with alcohol and a Q-Tip to remove deposits.
  6. Using a PETROLEUM BASED LUBRICANT, now you need to lubricate the O – Rings. This must be done BEFORE you INSTALL the VALVE CORES.
  7. INSTALL NEW VALVE CORES, using the VALVE CORE TOOL that comes with the kit.
  8. INSTALL NEW BUTTONS. REPLACE PIN on the side of the HOUSING.

REMOVAL & INSTALLATION (EURO – ROUND HEADED SYRINGE)

  1. DEPRESSURIZE dental station.
  2. REMOVE PIN from the side of the SYRINGE HOUSING.
  3. REMOVE BOTH BUTTONS and SPRINGS.
  4. CLEAN BOTH BUTTON HOLES with alcohol and a Q-Tip to remove deposits.
  5. LUBRICATE the O-RINGS.
  6. INSTALL NEW SPRINGS, smaller side UP.
  7. INSTALL NEW BUTTONS by PUSHING into ASSEMBLY all the way.
  8. REPLACE PIN on the side of the HOUSING.