INSTALLATION OF AIR/WATER SYRINGE KIT
Air/Water Syringe Repair Kit (General)
Product #: A113-019
Air/ Water Syringe Repair Kit (Euro)
Product #: A113-403
(Tools needed: Small Allen Key or a Hex Key)
Note: Petroleum Based Lubricant is needed for this repair
REMOVAL & INSTALLATION (GENERA – SQUARE HEADED SYRINGE)
- DEPRESSURIZE dental station.
- REMOVE PIN from the side of the SRINGE HOUSING by using a small PAPER CLIP or small ALLEN WRENCH.
- REMOVE BOTH BUTTONS by PULLING straight OUT.
- REMOVE BOTH VALVE CORES by UNSCREWING with tool that comes with the kit.
- CLEAN BOTH BUTTON HOLES with alcohol and a Q-Tip to remove deposits.
- Using a PETROLEUM BASED LUBRICANT, now you need to lubricate the O – Rings. This must be done BEFORE you INSTALL the VALVE CORES.
- INSTALL NEW VALVE CORES, using the VALVE CORE TOOL that comes with the kit.
- INSTALL NEW BUTTONS. REPLACE PIN on the side of the HOUSING.
REMOVAL & INSTALLATION (EURO – ROUND HEADED SYRINGE)
- DEPRESSURIZE dental station.
- REMOVE PIN from the side of the SYRINGE HOUSING.
- REMOVE BOTH BUTTONS and SPRINGS.
- CLEAN BOTH BUTTON HOLES with alcohol and a Q-Tip to remove deposits.
- LUBRICATE the O-RINGS.
- INSTALL NEW SPRINGS, smaller side UP.
- INSTALL NEW BUTTONS by PUSHING into ASSEMBLY all the way.
- REPLACE PIN on the side of the HOUSING.